GEOMAR moorings are typically equipped with instruments recording pressure, temperature, conductivity, dissolved oxygen and current velocity. Instruments with pressure, temperature, conductivity and oxygen sensors were calibrated in situ immediately prior to and after a mooring deployment period by attaching them to the CTD frame during CTDO casts. Correction terms were then developed from the difference between the sensor readings and the calibrated CTDO data during several minute long calibration stops. These correction terms were then applied to the full deployment periods. This ensured best data quality with recognition of potential sensor drifts and also allowed for the estimation of calibration and measurement errors (Hahn et al. 2014, Bittig et al. 2018, Berx et al. 2019).
instrument = Label:kpo_1170_02 File:kpo_1170_instr_02.nc Type:Teledyne RDI LongRanger 75 SN:17570instrument = Label:kpo_1170_03 File:kpo_1170_instr_03.nc Type:Seabird MicroCat 37-IM mit P SN:02717instrument = Label:kpo_1170_04 File:kpo_1170_instr_04.nc Type:Seabird MicroCat 37-SM SN:01682 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1170_05 File:kpo_1170_instr_05.nc Type:Nortek Aquadopp Seriell SN:P26209-27instrument = Label:kpo_1170_06 File:kpo_1170_instr_06.nc Type:Nortek Aquadopp Seriell SN:P26209-18instrument = Label:kpo_1170_07 File:kpo_1170_instr_07.nc Type:Seabird MicroCat 37-SM SN:01281 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1170_08 File:kpo_1170_instr_08.nc Type:Nortek Aquadopp Induktiv SN:P26209-02instrument = Label:kpo_1170_09 File:kpo_1170_instr_09.nc Type:Seabird MicroCat 37-IM SN:00954 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1170_10 File:kpo_1170_instr_10.nc Type:Seabird MicroCat 37-SM SN:01285 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1170_11 File:kpo_1170_instr_11.nc Type:Nortek Aquadopp Seriell SN:P26209-28instrument = Label:kpo_1170_12 File:kpo_1170_instr_12.nc Type:Seabird MicroCat 37-SM SN:01269 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailscomment = flag_meaningscomment = 0:unknowncomment = 1:good_datacomment = 2:probably_good_datacomment = 3:potentially_correctable_bad_datacomment = 4:bad_datacomment = 5:not_usedcomment = 6:not_usedcomment = 7:nominal_valuecomment = 8:interpolated_valuecomment = 9:missing_value