Figure S5 LSV varying compliance voltage

Figure S5: CV curves recorded in ambient air with two Potentiostats with a differing compliance voltage of 10 V and 12 V, respectively. Measurements were carried out on GDEs with a Pt loading of 100 µg cm.2 and a catalyst-coated area of 0.5 cm2 in 2 M HClO4. shows that the Potentiostat with the higher compliance voltage of 12 V allows a higher maximum current to flow through the cell. An interesting feature when comparing the curves recorded with the different Potentiostats is that the maximum current in the measurement with the Potentiostat with the lower compliance voltage is given as a plateau, while the Potentiostat with higher compliance voltage is stopping to record current once the compliance voltage is reached.

Identifier
Source https://tudatalib.ulb.tu-darmstadt.de/handle/tudatalib/3567
Metadata Access https://tudatalib.ulb.tu-darmstadt.de/oai/openairedata?verb=GetRecord&metadataPrefix=oai_datacite&identifier=oai:tudatalib.ulb.tu-darmstadt.de:tudatalib/3567
Provenance
Creator Schmitt, Nicolai; Schmidt, Mareike; Mueller, Jonathan E.; Schmidt, Lasse; Etzold, Bastian J.M.
Publisher TU Darmstadt
Contributor TU Darmstadt
Publication Year 2022
Rights Creative Commons Attribution 4.0; info:eu-repo/semantics/openAccess
OpenAccess true
Contact https://tudatalib.ulb.tu-darmstadt.de/page/contact
Representation
Resource Type Text
Format text/plain
Discipline Other